Biosafety Cabinet A2 by KLC

225

Product Introduction

1. User-friendly control software interface, simple operation, LCD screen, convenient and meet the customer’s needs. Real-time clock display, workspace temperature, and humidity display.

2. Preset filter failure and breakdown alarm technology. The control circuit can automatically initiate sound, light, and visual alarm prompt when the filters are near to or have reached their lifetime, the wind resistance increases, and the filter damaged.

3. With ULPA filter which can filter the 0.3um particle at 99.999% efficiency, and can filter the 0.1um particle at 99.995% efficiency.

4. It adopts asynchronous motor imported from Germany, it has the advantages of stable running, low noise, small power consumption, and low maintenance costs.

5. The front window is tempered glass, which with functions of good explosion-proof, shatterproof, and anti-ultraviolet. The equipment is designed with 10 degrees tilt angle, which conforms with the ergonomic principle, to reduce glare in the largest range.

6. The entire device around negative pressure air duct for left, right, and rear cabinet, so that the space between the work area and external environment will form air curtain and box type double isolation. In addition, we can make sure there‘s no leakage for our products because the work area is surrounded by negative pressure. Cabinet material is cold rolled steel plate in good strength. The working table-board and cabinet inside is stainless steel material, it has a good appearance and anti-corrosion. Supporting made by metal and electrostatic sprayed surface. Touch-type switch for control panel, it makes the equipment more wonderful and easy to operate.

Specification

ⅡType biosafety cabinet is safety equipment widely used in scientific research, education, clinical testing, and production in the field of microbiology, biomedicine, gene recombination, animal experiments, and biological products, etc, It applies a large number of unique designs and constructions in order to provide the best security protection for the environment, operators, and experimental items.

Construction characteristics

ⅡType biosafety cabinet is consists of the cabinet, frameless glass windows, SUS304 stainless steel table-board, front hand rack, built-in waterproof socket, caster, fan, air downflow filter, air exhaust filter, LED lamp, UV lamp, LCD screen controller, and alarm system. The cabinet has an attractive appearance which is made of a high-quality cold-rolled plate processed by laser cutting, CNC equipment, and electrostatic sprayed surface, it is matched with human mechanics’ 10 degrees tilt design, makes the operation more comfortable.

Product Specification

Item Biosafety cabinet

for single person A2

Small biosafety cabinet

for two-person A2

Big biosafety cabinet

for two-person A2

Type A2 – 30% air exhaust & 70% inner recycle
Model KLC-BSC1100ⅡA2-A KLC-BSC1500ⅡA2-A KLC-BSC1800ⅡA2-A
External size (W*L*H) 1100*750*2250 1500*750*2250 1870*750*2290
Internal size (W*L*H) 940*600*660 1350*600*660 1710*600*660
Filtration efficiency

(@0.3um)

99.999% 99.999% 99.999%
Average down flow 0.25~0.5m/s 0.25~0.5m/s 0.25~0.5m/s
Average in flow ≥0.5m/s ≥0.5m/s ≥0.5m/s
Light (Lux) ≥800 ≥800 ≥800
Noise (dB) ≤65 ≤65 ≤65
Vibration (μm) ≤5 ≤5 ≤5
Power (V/Hz) 220/50 220/50 220/50
Material Cabinet Q195/SUS201/SUS304

Table-board SUS201/SUS304

Biosafety Personnel safety: the number of colony of impact sampler≤10CFU/time

The colony number of slit sampler≤5CFU/time

Product safety: the colony number≤5CFU/time

Cross-contamination safety: the colony number≤2CFU/time

To know more, please check KLC.

Leave A Reply

Your email address will not be published.